단종 제품

BondMaster

The BondMasterTM 1000+ is a versatile, full multi-mode instrument, with Resonance, MIA (Mechanical Impedance Analysis), and Pitch/Catch mode with the addition of the new RF display mode. The BondMasterTM 1000+ is the only true multi-mode composite test instrument on the market. This offers the user the ability to select the method best suited to inspect the wide variety of composite materials used. Its high performance, lightweight, and rugged durability make it the choice for applications in the manufacturing, maintenance and repair of composite materials.
Weighing less than 4 lbs. (1.8 kg) with customer-interchangeable displays, the BondMaster 1000+ offers users the highest resolution available today. The availability of a color LCD or monochrome LCD for indoor or bright outdoor conditions or a Hi-Brite Electroluminescent (EL) display for normal to dark conditions provides the ultimate in flexibility and convenience. Its rugged, well designed housing, uncomplicated front panel, SmartKnobTM and built in PowerLinkTM technology, make it a truly revolutionary user-friendly hand held portable ultrasonic flaw detector.
The BondMaster 1000+ uses PowerLinkTM technology to configure the instrument to the technology in use when the sensor is connected. Built-in calibration modes assist the user in optimizing the test parameters. A variety of sensors in each of the test technologies is available.
The instrument uses a computer style lithium ion battery.

Features

• Single Li-Ion battery
• Lightweight, less than 4lbs. (1.8 Kg)
• Customer interchangeable displays:
- Hi-Brite Electroluminescent
- Monochrome Liquid Crystal
- Color Liquid Crystal
• VGA Output
• RS-232 output for printer or computer interface
• Alarm outputs
• Program storage
• Trace storage
• PowerLinkTM Technology - automatic probe recognition and instrument set-up
• Full Multi-Mode Capability:
- Pitch / Catch (Swept, Impulse, RF)
- Resonance
- MIA - Mechanical Impedance Analysis
• High voltage Pitch / Catch probes available

BondMaster 1000+ Specifications*


When a probe is connected to the BondMaster™, the instrument's mode and type of test are selected.
The modes of operation are Pitch / Catch (Swept, Impulse, RF), Resonance, and MIA.
Pitch/Catch Swept: Measures amplitude and phase changes using a swept frequency method to detect disbonds. Requires no couplant.
Pitch/Catch Impulse: Measures amplitude and phase changes using a short burst of energy to detect disbonds. Displays information in envelope or vector display format. Requires no couplant.
Pitch/Catch RF Impulse: Measures amplitude and phase changes using a short burst of energy to detect disbonds. Displays information in envelope or vector display format. An impedance display is available from the RF data. Requires no couplant.
Resonance: Detects disbonds by changes in phase and amplitude of probe resonance. Requires couplant.
Mechanical Impedance Analysis (MIA): Measures the stiffness and mass of the material under test. The output is measured in both amplitude and phase. Requires no couplant.

Probes
All BondMaster™ 1000 probes include PowerLink™ functionality. S-PC, S-MP, and S-PR probes are compatible.

Technical Specifications
FrequencyRange: 250 Hz to 1.5 MHz. Specific test modes have limitations within this range.
Gain: -10 to 50.0 dB
Analog Output Update: Rate continuous in MIA and resonance modes. Data available at repetition rate in all pitch/catch modes
Alarm Box: The instrument allows for any size alarm box. The box can be defined and located anywhere on the screen. An adjustable vertical amplitude alarm operates in the RF and Impulse setup modes.
Alarm Logic: Positive or negative alarm gate
Alarm Output: 0 V to +3 V HC logic output, switchable audible alarm and front panel annunciator, alarm indicator on probe is standard.
Clock / Calendar: Time and date are stored and printed with each waveform.
Languages: Menus may be displayed in English, Spanish, French, or German.
RS-232 Interface: Screen printout and computer interface. Baud rate is adjustable from 2,400 to 19,200 baud (9,600 is default).
Screen Storage: Up to 20 screens can be stored.
Program Storage: Up to 100 instrument setups

Inputs / Outputs
Probe Connector: 11-pin Fisher
Analog Outputs: Signals: +/- 5 volt, adjustable offset, not affected by position controls or ZOOM function.

General
Dimensions: 9.5" L x 5.5" H x 3.6" D (215 mm x 165 mm x 92 mm)
Weight: 3 to 6 lbs. (1.3 - 2.7 kg) depending on configuration
Display: Customer-interchangeable QVGA displays (320 x 240 pixels) color or monochrome LCD, Hi-Brite Electroluminescent
Operating Temperature: -4o to 140o F (-20o to 60o C)
Storage Temperature: -40o to 176o F (-40o to 80o C)
Humidity: 95% +/- 5%
Classification: Based on Class 2 specifications from the MIL-PRF-28800F Handbook
Altitude: Maximum operating and non-operating altitude- 15,000 ft. (4,600 m)
Hazardous Area Operation: Safe operation as defined by Class I, Division 2, Group D, as found in the National Fire Association Code (NFPA 70), Section 500, and tested using MIL-STD-810F, Method 511.4, Procedure 1
Power
Power: 7-pin connector to charge the internal batteries and operate the instrument from AC power
Power Requirements: 85 to 240 volts, 50-60 Hz mains. External holder charges batteries outside the instrument. Charging time is typically 4 hours.
Low Battery Protection: Display bar graph "gas gauge" indicates approximate operating time.
BatteryOperating Time: 6 to 8 hours (nominal, depending on configuration)

*All specifications are subject to change without notice.