Soil samples collected from key locations from the grounds of a 200 year old building in New England illustrate expected variations in Lead (Pb) content. A DELTA analyzer configured in SmartShot Soil analysis mode was used to analyze the samples. Spectral data is shown using the DELTA PC Software. |
A DELTA handheld XRF analyzer, configured in SmartShot soil analysis mode, was used to run repeat measurements on a reference soil sample, to demonstrate repeatability. As shown below, the DELTA PC Software can be used to overlay spectra from different readings for direct comparison and analysis. |
A reference sample spiked with approximately 500ppm Pb (Innov-X sample 408) was used for these repeat readings. Each reading was analyzed for 60 seconds. +/- error shown is 2-sigma, 95% confidence. |
| |||||||||||||||||||||
Die DELTA Handgeräte zur Röntgenfluoreszenzanalyse (RFA) prüfen gefährlich hohe Pegel der wichtigsten Schadstoffe und gesundheitsschädlichen Metalle in Sekunden entsprechend RCRA. Sie werden eingesetzt zur Prüfung von Boden- und Sedimentproben nach EPA 6200 und anderen Standardarbeitsanweisungen, zur Analyse von Filtern nach der NIOSH- und OSHA-Methode, und zum Messen des Gehalts an Pb, Cd, Cr, As, Hg und anderen Stoffen an Flächen und in Staubproben.Copyright 2011 OLYMPUS CORPORATION, Alle Rechte vorbehalten. Nutzungsbedingungen | Datenschutzerklärung