Normal Incidence Shear Wave Transducers


Application

This document summarizes important information for users of normal incidence shear wave transducers such as the V150-V157 and V220-V222 series. These transducers are most commonly used with ultrasonic thickness gages, flaw detectors, or pulser/receivers for measurement of shear wave velocity in test materials, often in connection with the calculation of elastic modulus as described in the application note Elastic Modulus Measurement. 

Normal Incidence Shear Wave Transducers

Background

Normal incidence shear wave transducers are single element contact transducers that introduce shear waves directly into the test piece without the use of refractive mode conversion. By definition, these waves propagate perpendicular to the test surface while particle motion is parallel to the surface. The transducers use piezoelectric elements cut perpendicular to the electrical poling direction so that, when pulsed, they vibrate in shear (transverse) mode rather than longitudinal (compressional) mode when an excitation voltage is applied, as seen in the concept drawing below. 

Longitudinal wave element
Longitudinal wave element
Shear wave element
Shear wave element

Because of their specialized nature, shear wave transducers require attention to the factors noted below for good response.

(a) Excitation pulse limit

Shear wave transducers should always be driven with the lowest excitation voltage possible in a given test, preferably 100 volts. When too much power is delivered to the transducer over time, the element can actually be re-poled to a longitudinal mode element. This effect can occur with long-term use of the 300–400 V pulsing available in most ultrasonic flaw detectors. Higher frequency elements (5 MHz and up) are most susceptible to this re-poling because they are thinner. There is no way to reverse this effect, and, when damaged this way, the re-poled transducer essentially becomes a longitudinal wave transducer. 

(b) Coupling

A high viscosity shear wave couplant, such as Olympus SWC-2 (Q7700010), must be used to couple a normal incidence shear wave into the test piece. Standard ultrasonic couplants will not work because they are liquids or gels. A basic property of a liquid is that it does not support a shear stress. Thus, low and medium viscosity liquids, such as conventional ultrasonic couplants, will not transmit shear waves. For best results, it is important to use a very thin layer of the SWC-2 high viscosity couplant with firm coupling pressure. The suggested procedure is to put a very small amount of couplant on the transducer and spread it into a thin layer with a razor blade or straight edge, then couple onto the test piece and further wring out the couplant by rotating the transducer. You will typically see the echo amplitude increase as the couplant layer becomes progressively thinner. 

(c) Polarization direction

The direction of polarization of the shear wave (the axis of particle movement) in Olympus shear wave transducers is nominally in line with the right angle connector in standard RM and RB style cases. Shear wave transducers with SB or SM connectors have a line engraved on the case marking the polarization axis. In anisotropic materials, it is common that pulse transit time and amplitude will vary as the transducer is rotated on the surface of the test piece, which changes the orientation of particle movement with respect to directionally varying mechanical properties of the material. 

(d) Material attenuation

Shear wave propagation is typically good in common engineering metals and ceramics unless a coarse grain structure is present. However, shear wave attenuation is usually extremely high in flexible materials such as rubber and soft plastics, and, thus, there will typically be no usable shear wave echo, even with proper couplant. Hard plastics, like acrylics, as well as structural composites usually do transmit usable shear waves at low ultrasonic test frequencies, but care should be taken with transducer selection and instrument setup. 

(e) Longitudinal wave artifacts

All normal incidence shear wave elements generate some underlying longitudinal wave energy as well. Typically, this longitudinal component is at least 30 dB lower than the shear signal, however, in materials with very high shear attenuation and lower L-wave attenuation (like flexible plastics) or in situations where non-viscous couplants are used, the shear component may be highly attenuated while some L-wave energy remains and is seen as the primary waveform in the display. This phenomenon can also occur when shear wave transducers have been damaged by excessively high excitation voltages as noted in section (a) above. 

For further information, please contact us.

Olympus IMS
应用所使用的产品

EPOCH 6LT便携式超声探伤仪经过优化可使操作人员单手完成操作。在需要借助绳索攀爬进行的检测以及对便携性能有很高要求的应用中,这款探伤仪可表现出卓越的性能。探伤仪的重量很轻,机身设计符合人体工程学的要求,操作人员单手即可稳固地持握仪器,在需要借助绳索攀爬进行的检测应用中,也可以将仪器绑在腿上,完成无需用手持握仪器的操作。
超声探头可以用于各种各样的应用,如:缺陷探伤、厚度测量、材料研究及医学诊断。具有各种晶片直径、频率和连接器类型的不同类型超声探头达5000多种。
EPOCH 650仪器是一款常规超声探伤仪,具有优质的检测性能和广泛的适用性,可用于各种各样的应用。这款简单直观、坚固耐用的仪器是广受欢迎的EPOCH 600探伤仪的延续产品,在原有的仪器中添加了更多的功能。
手持式45MG是一款配置有各种测量功能和软件选项的超声测厚仪。这款独特的仪器与所有Olympus双晶和单晶探头探头兼容,是一款集所有解决方案于一机的创新型仪器,可用于几乎所有测厚应用。
38DL PLUS是一款高级超声测厚仪。这款仪器可使用双晶探头对内部腐蚀的部件进行检测。其性能包含THRU-COAT®(穿透涂层)和回波到回波。还可以使用单晶探头对薄材料、极厚材料以及多层材料进行非常精确的厚度测量。
Sorry, this page is not available in your country
Let us know what you're looking for by filling out the form below.

This site uses cookies to enhance performance, analyze traffic, and for ads measurement purposes. If you do not change your web settings, cookies will continue to be used on this website. To learn more about how we use cookies on this website, and how you can restrict our use of cookies, please review our Cookie Policy.

OK