Vanta™ iX Specifications
Dimensions (W × H × D) | 10 cm × 7.9 cm × 26.6 cm (3.9 in. × 3.1 in. × 10.5 in.) |
Weight | 2.4 kg (5.29 lb) |
Excitation Source |
X-ray tube: Rh or W anode (application optimized) 5–200 μA
MR model: 8–50 keV (4 W max) CW model: 8–40 keV (4 W max) |
Primary Beam Filtration | Eight filter positions automatically selected per beam per method |
Detector |
MR model: Large-area silicon drift detector
CW model: Standard silicon drift detector |
Power | Power over Ethernet (PoE) or 18 V AC power adaptor |
Elemental Range |
Method dependent:
MR model: Mg–U CW model: Ti–U (with standard window and calibration) |
Pressure Correction | Built-in barometer for automatic altitude and air pressure correction |
IP Rating | IP54 |
Operating Environment |
Temperature range: –10 °C to 50 °C (14 °F to 122 °F) under continuous duty cycle
Humidity: 10% to 90% relative humidity, non-condensing |
Operating System | Linux |
Application Software | Olympus proprietary data acquisition and processing package |