Application

This document summarizes important information for users of normal incidence shear wave transducers such as the V150-V157 and V220-V222 series. These transducers are most commonly used with ultrasonic thickness gages, flaw detectors, or pulser/receivers for measurement of shear wave velocity in test materials, often in connection with the calculation of elastic modulus as described in the application note Elastic Modulus Measurement. 

Normal Incidence Shear Wave Transducers

Background

Normal incidence shear wave transducers are single element contact transducers that introduce shear waves directly into the test piece without the use of refractive mode conversion. By definition, these waves propagate perpendicular to the test surface while particle motion is parallel to the surface. The transducers use piezoelectric elements cut perpendicular to the electrical poling direction so that, when pulsed, they vibrate in shear (transverse) mode rather than longitudinal (compressional) mode when an excitation voltage is applied, as seen in the concept drawing below. 

Longitudinal wave element
Longitudinal wave element
Shear wave element
Shear wave element

Because of their specialized nature, shear wave transducers require attention to the factors noted below for good response.

(a) Excitation pulse limit

Shear wave transducers should always be driven with the lowest excitation voltage possible in a given test, preferably 100 volts. When too much power is delivered to the transducer over time, the element can actually be re-poled to a longitudinal mode element. This effect can occur with long-term use of the 300–400 V pulsing available in most ultrasonic flaw detectors. Higher frequency elements (5 MHz and up) are most susceptible to this re-poling because they are thinner. There is no way to reverse this effect, and, when damaged this way, the re-poled transducer essentially becomes a longitudinal wave transducer. 

(b) Coupling

A high viscosity shear wave couplant, such as Olympus SWC-2 (Q7700010), must be used to couple a normal incidence shear wave into the test piece. Standard ultrasonic couplants will not work because they are liquids or gels. A basic property of a liquid is that it does not support a shear stress. Thus, low and medium viscosity liquids, such as conventional ultrasonic couplants, will not transmit shear waves. For best results, it is important to use a very thin layer of the SWC-2 high viscosity couplant with firm coupling pressure. The suggested procedure is to put a very small amount of couplant on the transducer and spread it into a thin layer with a razor blade or straight edge, then couple onto the test piece and further wring out the couplant by rotating the transducer. You will typically see the echo amplitude increase as the couplant layer becomes progressively thinner. 

(c) Polarization direction

The direction of polarization of the shear wave (the axis of particle movement) in Olympus shear wave transducers is nominally in line with the right angle connector in standard RM and RB style cases. Shear wave transducers with SB or SM connectors have a line engraved on the case marking the polarization axis. In anisotropic materials, it is common that pulse transit time and amplitude will vary as the transducer is rotated on the surface of the test piece, which changes the orientation of particle movement with respect to directionally varying mechanical properties of the material. 

(d) Material attenuation

Shear wave propagation is typically good in common engineering metals and ceramics unless a coarse grain structure is present. However, shear wave attenuation is usually extremely high in flexible materials such as rubber and soft plastics, and, thus, there will typically be no usable shear wave echo, even with proper couplant. Hard plastics, like acrylics, as well as structural composites usually do transmit usable shear waves at low ultrasonic test frequencies, but care should be taken with transducer selection and instrument setup. 

(e) Longitudinal wave artifacts

All normal incidence shear wave elements generate some underlying longitudinal wave energy as well. Typically, this longitudinal component is at least 30 dB lower than the shear signal, however, in materials with very high shear attenuation and lower L-wave attenuation (like flexible plastics) or in situations where non-viscous couplants are used, the shear component may be highly attenuated while some L-wave energy remains and is seen as the primary waveform in the display. This phenomenon can also occur when shear wave transducers have been damaged by excessively high excitation voltages as noted in section (a) above. 

For further information, please contact us.

Olympus IMS

이 애플리케이션에 사용되는 제품

72DL PLUS™ 고급 초음파 두께 측정기는 사용이 간편한 휴대용 장치로 빠르고 정밀하게 두께를 측정합니다.최대 125MHz의 단일 요소 탐촉자와 호환되는 이 혁신적인 두께 측정 도구는 다층 도장, 코팅, 플라스틱과 같은 초박막 소재의 두께의 측정에 매우 적합합니다.최대 6층의 두께를 동시에 표시할 수 있습니다.

초음파 탐촉자는 결함 감지, 두께 측정, 소재 연구 및 의료 진단과 같은 다양한 응용 분야에 사용됩니다. 5000개 이상의 초음파 탐촉자 유형은 다양한 스타일, 요소 직경, 주파수 및 커넥터 스타일로 제공됩니다.
EPOCH 6LT 휴대용 초음파 결함 탐상기는 한 손 작업에 최적화되어 있으며 로프 접근 및 고소 작업이 필요한 고휴대성 응용 분야에서 탁월한 성능을 제공합니다. 가볍고 인체 공학적 디자인으로 사용자의 손에 단단히 고정시키나 로프 접근이 필요한 응용 분야에서는 다리에 묶을 수 있습니다.
EPOCH 650은 다양한 응용 분야에 대한 우수한 검사 성능과 유용성을 갖춘 재래식 초음파 결함 검출기입니다. 이 직관적이고 견고한 장비는 인기 있는 EPOCH 600 결함 탐상기 후속작이며 추가 기능을 갖추고 있습니다.
45MG 고급 초음파 두께 측정기에는 표준 측정 기능은 물론 다양한 소프트웨어 옵션이 있습니다.이 독특한 두께 측정 도구는 당사의 이중 요소 및 단일 요소 두께 측정 탐촉자와 호환됩니다.

다용도로 활용 가능한 38DL PLUS™ 측정기를 이중 요소 탐촉자와 함께 사용하면 부식된 파이프의 두께를 측정할 수 있으며, 단일 요소 탐촉자를 사용하면 박층 또는 다층 소재의 두께를 매우 정확하게 측정할 수 있습니다.

죄송합니다. 이 페이지는 해당 국가에서 사용할 수 없습니다.
아래 양식을 작성하여 원하는 내용을 알려주십시오.
죄송합니다. 이 페이지는 해당 국가에서 사용할 수 없습니다.