Evident LogoOlympus Logo
Documentation
Notes d’application
Retour aux ressources

Normal Incidence Shear Wave Transducers


Application

This document summarizes important information for users of normal incidence shear wave transducers such as the V150-V157 and V220-V222 series. These transducers are most commonly used with ultrasonic thickness gages, flaw detectors, or pulser/receivers for measurement of shear wave velocity in test materials, often in connection with the calculation of elastic modulus as described in the application note Elastic Modulus Measurement. 

Normal Incidence Shear Wave Transducers

Background

Normal incidence shear wave transducers are single element contact transducers that introduce shear waves directly into the test piece without the use of refractive mode conversion. By definition, these waves propagate perpendicular to the test surface while particle motion is parallel to the surface. The transducers use piezoelectric elements cut perpendicular to the electrical poling direction so that, when pulsed, they vibrate in shear (transverse) mode rather than longitudinal (compressional) mode when an excitation voltage is applied, as seen in the concept drawing below. 

Longitudinal wave element
Longitudinal wave element
Shear wave element
Shear wave element

Because of their specialized nature, shear wave transducers require attention to the factors noted below for good response.

(a) Excitation pulse limit

Shear wave transducers should always be driven with the lowest excitation voltage possible in a given test, preferably 100 volts. When too much power is delivered to the transducer over time, the element can actually be re-poled to a longitudinal mode element. This effect can occur with long-term use of the 300–400 V pulsing available in most ultrasonic flaw detectors. Higher frequency elements (5 MHz and up) are most susceptible to this re-poling because they are thinner. There is no way to reverse this effect, and, when damaged this way, the re-poled transducer essentially becomes a longitudinal wave transducer. 

(b) Coupling

A high viscosity shear wave couplant, such as Olympus SWC-2 (Q7700010), must be used to couple a normal incidence shear wave into the test piece. Standard ultrasonic couplants will not work because they are liquids or gels. A basic property of a liquid is that it does not support a shear stress. Thus, low and medium viscosity liquids, such as conventional ultrasonic couplants, will not transmit shear waves. For best results, it is important to use a very thin layer of the SWC-2 high viscosity couplant with firm coupling pressure. The suggested procedure is to put a very small amount of couplant on the transducer and spread it into a thin layer with a razor blade or straight edge, then couple onto the test piece and further wring out the couplant by rotating the transducer. You will typically see the echo amplitude increase as the couplant layer becomes progressively thinner. 

(c) Polarization direction

The direction of polarization of the shear wave (the axis of particle movement) in Olympus shear wave transducers is nominally in line with the right angle connector in standard RM and RB style cases. Shear wave transducers with SB or SM connectors have a line engraved on the case marking the polarization axis. In anisotropic materials, it is common that pulse transit time and amplitude will vary as the transducer is rotated on the surface of the test piece, which changes the orientation of particle movement with respect to directionally varying mechanical properties of the material. 

(d) Material attenuation

Shear wave propagation is typically good in common engineering metals and ceramics unless a coarse grain structure is present. However, shear wave attenuation is usually extremely high in flexible materials such as rubber and soft plastics, and, thus, there will typically be no usable shear wave echo, even with proper couplant. Hard plastics, like acrylics, as well as structural composites usually do transmit usable shear waves at low ultrasonic test frequencies, but care should be taken with transducer selection and instrument setup. 

(e) Longitudinal wave artifacts

All normal incidence shear wave elements generate some underlying longitudinal wave energy as well. Typically, this longitudinal component is at least 30 dB lower than the shear signal, however, in materials with very high shear attenuation and lower L-wave attenuation (like flexible plastics) or in situations where non-viscous couplants are used, the shear component may be highly attenuated while some L-wave energy remains and is seen as the primary waveform in the display. This phenomenon can also occur when shear wave transducers have been damaged by excessively high excitation voltages as noted in section (a) above. 

For further information, please contact us.

Olympus IMS

Produits utilisés pour cette application

Portable et facile à utiliser, le mesureur d’épaisseur à ultrasons 72DL PLUS™ peut fournir très rapidement des mesures d’épaisseur précises. Compatible avec les sondes monoéléments d’une fréquence allant jusqu’à 125 MHz, cet appareil novateur est parfaitement adapté à la mesure de l’épaisseur des matériaux ultrafins, notamment les peintures multicouches, les revêtements et le plastique. Le logiciel de mesure multicouche peut afficher simultanément l’épaisseur de six couches distinctes.

Les sondes à ultrasons sont utilisées pour des applications aussi diverses que la recherche de défauts, la mesure d’épaisseur, la recherche sur les matériaux et les diagnostics médicaux. Il existe plus de 5000 types de sondes à ultrasons de différents styles offrant différents diamètres d’éléments, fréquences et connecteurs.
L’appareil portable de recherche de défauts par ultrasons EPOCH 6LT est optimisé pour le fonctionnement à une seule main et offre une excellente performance pour les applications nécessitant un accès par cordes requérant une très grande portabilité. Ergonomique et léger, l’appareil est parfaitement adapté à la main de l’utilisateur et peut être attaché à la jambe pour faciliter les inspections nécessitant un accès par cordes.
L’EPOCH 650 est un appareil de recherche de défauts par ultrasons conventionnels très performant adapté à un grand nombre d’applications. Cet appareil intuitif et robuste remplace le populaire appareil de recherche de défauts EPOCH 600. Il est aussi doté de fonctionnalités supplémentaires.
Le 45MG est un mesureur d’épaisseur à ultrasons de pointe équipé de série de multiples fonctions de mesure et d’options logicielles. Cet outil de mesure d’épaisseur unique est compatible avec notre gamme complète de sondes de mesure d’épaisseur monoéléments et à émission-réception séparées.
Polyvalent, le mesureur d’épaisseur 38DL PLUS peut être combiné à des sondes à émission-réception séparées pour la mesure de l’épaisseur de tuyaux corrodés, ou à une sonde monoélément pour la mesure très précise de l’épaisseur de matériaux minces ou multicouches.
Sorry, this page is not available in your country
Let us know what you're looking for by filling out the form below.
Sorry, this page is not available in your country