Thickness gaging high-temperature

Crude oil is an important source of energy, but using it in its natural form is inefficient and possibly dangerous. Because of this, refineries turn crude oil into finished petroleum products.

The refining process starts with heating the crude oil and putting it into a distillation column (also known as a still). In the still, diesel, gas, and other products boil off and are recovered at different temperatures. The distillation temperature for petroleum products ranges from as low as 30 °C (85 °F) to greater than 565 °C (1,050 °F). These high surface temperatures make inspecting refinery equipment challenging.

Challenge: Too Hot to Handle

Periodic thickness measurements using ultrasonic testing (UT) are needed to monitor for corrosion and other defects, which can cause equipment failure to potentially shut down the refinery’s operation. Since the refining process runs continuously, the machinery can’t be cooled down without stopping the entire process. But high surface temperatures can easily destroy a thickness gage transducer; many can only tolerate temperatures up to about 50 °C, or 125 °F.

Solution: Transducers That Can Take the Heat

Unlike traditional UT transducers, dual element transducers have delay material that acts as insulation between the active transducer element and the hot test surface. They’re frequently recommended for high-temperature applications because some—such as the D790—can be used on surfaces as hot as 500 °C (930 °F) with intermittent contact (5 seconds or less).

Corrosion gaging duals with temperature specifications.
Corrosion gaging duals with temperature specifications.

Challenge: Changing Sound Velocity

Accurate UT thickness measurements depend, in part, on the velocity of sound in the test material. As a material’s temperature increases, the sound velocity decreases, complicating the accuracy of the thickness readings. According to ASTM E797-95, the sound velocity of carbon steel decreases by “about 1% per 55 °C (100 °F)” increase in temperature.

To get around this problem, some inspectors will calibrate their flaw detector or thickness gage on a hot test block. To heat the block, they’d hold it to the pipe or tank being tested. However, this process is slow, and the temperature of the block wouldn’t exactly match the test material, causing imprecise measurements.

An alternative method is to take the thickness measurements and then manually calculate a correction based on the part’s thickness and temperature. This technique is time consuming, and the operator is not able to see the true thickness reading during the inspection.

Corrosion gaging duals with temperature specifications.

Solution: Automatic Temperature Compensation

Our predecessor, Panametrics NDT, pioneered a solution to this challenge nearly 15 years ago. Today, the 38DL PLUS® ultrasonic thickness gage includes Temperature Compensation software (as a standard option) that enables you to calibrate the gage at room temperature and then simply input the elevated temperature of the material being tested. The gage will automatically compensate for the velocity change and display the temperature-compensated thickness. The software is easy to use and requires little effort to set up, so you can calibrate quickly and receive real-time temperature-adjusted measurements. It’s that simple.

Related Content

8 Tips to Improve Your High-Temperature Corrosion Thickness Measurements

High-Temperature Ultrasonic Testing

Ultrasonic Thickness Measurements of Hot Metals

Introduction to Thickness Gaging

Product Marketing Specialist, Ultrasonic Flaw Detectors

August 28, 2018
InSight Blog Sign-up

By clicking subscribe you are agreeing to our privacy policy which can be found here.

Sorry, this page is not available in your country

Let us know what you're looking for by filling out the form below.

This site uses cookies to enhance performance, analyze traffic, and for ads measurement purposes. If you do not change your web settings, cookies will continue to be used on this website. To learn more about how we use cookies on this website, and how you can restrict our use of cookies, please review our Cookie Policy.

OK