Application

This document summarizes important information for users of normal incidence shear wave transducers such as the V150-V157 and V220-V222 series. These transducers are most commonly used with ultrasonic thickness gages, flaw detectors, or pulser/receivers for measurement of shear wave velocity in test materials, often in connection with the calculation of elastic modulus as described in the application note Elastic Modulus Measurement. 

Normal Incidence Shear Wave Transducers

Background

Normal incidence shear wave transducers are single element contact transducers that introduce shear waves directly into the test piece without the use of refractive mode conversion. By definition, these waves propagate perpendicular to the test surface while particle motion is parallel to the surface. The transducers use piezoelectric elements cut perpendicular to the electrical poling direction so that, when pulsed, they vibrate in shear (transverse) mode rather than longitudinal (compressional) mode when an excitation voltage is applied, as seen in the concept drawing below. 

Longitudinal wave element
Longitudinal wave element
Shear wave element
Shear wave element

Because of their specialized nature, shear wave transducers require attention to the factors noted below for good response.

(a) Excitation pulse limit

Shear wave transducers should always be driven with the lowest excitation voltage possible in a given test, preferably 100 volts. When too much power is delivered to the transducer over time, the element can actually be re-poled to a longitudinal mode element. This effect can occur with long-term use of the 300–400 V pulsing available in most ultrasonic flaw detectors. Higher frequency elements (5 MHz and up) are most susceptible to this re-poling because they are thinner. There is no way to reverse this effect, and, when damaged this way, the re-poled transducer essentially becomes a longitudinal wave transducer. 

(b) Coupling

A high viscosity shear wave couplant, such as Olympus SWC-2 (Q7700010), must be used to couple a normal incidence shear wave into the test piece. Standard ultrasonic couplants will not work because they are liquids or gels. A basic property of a liquid is that it does not support a shear stress. Thus, low and medium viscosity liquids, such as conventional ultrasonic couplants, will not transmit shear waves. For best results, it is important to use a very thin layer of the SWC-2 high viscosity couplant with firm coupling pressure. The suggested procedure is to put a very small amount of couplant on the transducer and spread it into a thin layer with a razor blade or straight edge, then couple onto the test piece and further wring out the couplant by rotating the transducer. You will typically see the echo amplitude increase as the couplant layer becomes progressively thinner. 

(c) Polarization direction

The direction of polarization of the shear wave (the axis of particle movement) in Olympus shear wave transducers is nominally in line with the right angle connector in standard RM and RB style cases. Shear wave transducers with SB or SM connectors have a line engraved on the case marking the polarization axis. In anisotropic materials, it is common that pulse transit time and amplitude will vary as the transducer is rotated on the surface of the test piece, which changes the orientation of particle movement with respect to directionally varying mechanical properties of the material. 

(d) Material attenuation

Shear wave propagation is typically good in common engineering metals and ceramics unless a coarse grain structure is present. However, shear wave attenuation is usually extremely high in flexible materials such as rubber and soft plastics, and, thus, there will typically be no usable shear wave echo, even with proper couplant. Hard plastics, like acrylics, as well as structural composites usually do transmit usable shear waves at low ultrasonic test frequencies, but care should be taken with transducer selection and instrument setup. 

(e) Longitudinal wave artifacts

All normal incidence shear wave elements generate some underlying longitudinal wave energy as well. Typically, this longitudinal component is at least 30 dB lower than the shear signal, however, in materials with very high shear attenuation and lower L-wave attenuation (like flexible plastics) or in situations where non-viscous couplants are used, the shear component may be highly attenuated while some L-wave energy remains and is seen as the primary waveform in the display. This phenomenon can also occur when shear wave transducers have been damaged by excessively high excitation voltages as noted in section (a) above. 

For further information, please contact us.

Olympus IMS

Verwendete Produkte

Das hochmoderne 72DL PLUS Ultraschall-Dickenmessgerät liefert präzise Dickenmessungen mit hoher Geschwindigkeit in einem portablen, bedienerfreundlichen Gerät. Dieses innovative Dickenmessgerät ist mit Einzelschwinger-Messköpfen bis zu 125 MHz kompatibel und eignet sich zur Messung der Dicke von sehr dünnen Werkstoffen, einschließlich mehrschichtiger Anstrichfarbe, Beschichtungen und Kunststoffen. Es kann die Dicke von bis zu 6 Schichten gleichzeitig anzeigen.

Ultraschallköpfe werden in den verschiedensten Anwendungsbereichen eingesetzt, zur Fehlerprüfung, zum Messen der Dicke, zur Materialforschung und medizinischen Diagnose. Wir stellen über 5000 Ultraschallköpfe in vielen verschiedenen Arten, Elementdurchmessern, Frequenzen und Anschlusstypen her.
Die tragbaren Ultraschallprüfgeräte EPOCH 6LT sind für die Bedienung mit einer Hand optimiert und bieten eine hervorragende Leistung bei Anwendungen in Verbindung mit der Seilzugangstechnik und bei Anwendungen, die ein hohes Maß an Beweglichkeit und Flexibilität erfordern. Die Geräte mit ihrem leichten und ergonomischen Design liegen sicher in der Hand oder können in Verbindung mit der Seilzugangstechnik am Bein des Prüfers befestigt werden.
Die leistungsstarken EPOCH 650 Prüfgeräte für konventionellen Ultraschall können für zahlreiche Anwendungen eingesetzt werden. Diese intuitiven und robusten Prüfgeräte mit zusätzlichen Funktionen sind die Fortführung der bekannten Prüfgeräte EPOCH 600.

Der 45MG ist ein hochmodernes Ultraschall-Dickenmessgerät mit Standard-Messfunktionen und Softwareoptionen. Dieses einzigartige Dickenmessgerät ist mit allen unseren Sender-Empfänger-Messköpfen und Einzelschwingermessköpfen kompatibel.

Das vielseitig einsetzbare 38DL PLUS Messgerät kann mit Sender-Empfänger-Messköpfen zur Messung korrodierter Rohre bis hin zur sehr präzisen Dickenmessung von dünnen oder mehrschichtigen Materialien mit einem Einzelschwingermesskopf verwendet werden.
Sorry, this page is not available in your country
Let us know what you're looking for by filling out the form below.
Sorry, this page is not available in your country