1 00:00:06,953 --> 00:00:10,211 This video will demonstrate how to configure the NORTEC 600 2 00:00:10,213 --> 00:00:14,065 for surface and subsurface inspections. 3 00:00:14,066 --> 00:00:17,513 The NORTEC 600 can be configured in various ways. 4 00:00:17,515 --> 00:00:19,480 The following method is the fastest way 5 00:00:19,481 --> 00:00:21,118 to obtain good signals 6 00:00:21,120 --> 00:00:23,353 and consistent results. 7 00:00:23,355 --> 00:00:25,883 Connect the probe to the Nortec 600 8 00:00:25,885 --> 00:00:28,038 and turn the instrument on. 9 00:00:28,040 --> 00:00:29,928 With the latest software revision, 10 00:00:29,930 --> 00:00:33,248 the Application Selection menu will pop up automatically. 11 00:00:33,250 --> 00:00:36,950 Select Surface Cracks and press the checkmark key. 12 00:00:36,951 --> 00:00:38,998 Enter the correct test frequency. 13 00:00:39,000 --> 00:00:44,560 For our purposes we will use a default frequency of 500kHz. 14 00:00:44,561 --> 00:00:47,356 Press DISP and move the crosshair 15 00:00:47,358 --> 00:00:50,183 using the C key which is the position option. 16 00:00:50,185 --> 00:00:54,363 We will set the position to bottom center by using the knob. 17 00:00:54,365 --> 00:00:57,231 There are five commonly used industry presets 18 00:00:57,233 --> 00:00:59,698 to facilitate the placement of the crosshair. 19 00:00:59,700 --> 00:01:03,983 A sixth preset is available for user customization. 20 00:01:03,985 --> 00:01:08,061 Place the probe on a defect free area of the standard. 21 00:01:08,063 --> 00:01:10,255 Press and hold the null key. 22 00:01:10,256 --> 00:01:12,975 This will activate the auto lift off function. 23 00:01:12,976 --> 00:01:15,976 When you hear the beep, lift the probe. 24 00:01:15,978 --> 00:01:19,675 The cursor should appear on the horizontal axis to the far left. 25 00:01:19,676 --> 00:01:22,188 If it does not, repeat the step. 26 00:01:22,190 --> 00:01:24,600 Now put the probe on a defect free area 27 00:01:24,601 --> 00:01:26,861 and press null once. 28 00:01:26,863 --> 00:01:30,385 Scan the sample and press the freeze key. 29 00:01:30,386 --> 00:01:34,466 Once in Freeze mode, an F should appear below the battery indicator. 30 00:01:34,468 --> 00:01:36,191 Freeze mode enables the user 31 00:01:36,193 --> 00:01:38,516 to calibrate the signal using both hands. 32 00:01:38,518 --> 00:01:41,405 Press the main key to fine tune the horizontal angle 33 00:01:41,406 --> 00:01:44,335 and try to set it as flat as possible. 34 00:01:44,336 --> 00:01:46,313 Use the knob to adjust the angle 35 00:01:46,315 --> 00:01:49,681 and the checkmark key to go from coarse to fine tuning. 36 00:01:49,683 --> 00:01:52,106 Pressing the gain key once will enable the user 37 00:01:52,108 --> 00:01:55,891 to modify both the Horizontal and Vertical gains together. 38 00:01:55,893 --> 00:01:59,696 Press the Gain key once more to modify the H Gain. 39 00:01:59,698 --> 00:02:01,800 Reduce the gain so that the highest signal 40 00:02:01,801 --> 00:02:04,271 covers about 3 horizontal divisions. 41 00:02:04,273 --> 00:02:07,236 Press the gain key again to modify the V gain. 42 00:02:07,238 --> 00:02:09,081 Increase the gain so that the highest signal 43 00:02:09,083 --> 00:02:11,690 covers about 90% of the screen height. 44 00:02:11,691 --> 00:02:14,316 Once completed press freeze. 45 00:02:14,318 --> 00:02:17,300 Press Full Next to enter full screen mode. 46 00:02:17,301 --> 00:02:21,078 Place the probe on a defect free area and press null. 47 00:02:21,080 --> 00:02:23,130 Rescan the standard, 48 00:02:23,131 --> 00:02:28,883 The newly acquired scan is now calibrated. 49 00:02:28,885 --> 00:02:32,988 Subsurface Cracks - Generic Setup on Aluminum. 50 00:02:32,990 --> 00:02:34,995 For subsurface crack inspection 51 00:02:34,996 --> 00:02:39,901 connect an appropriate probe to the Nortec 600. 52 00:02:39,903 --> 00:02:43,648 Press the advanced setup key, then application selection. 53 00:02:43,650 --> 00:02:46,120 Select subsurface from the menu using the knob 54 00:02:46,121 --> 00:02:48,820 and press the checkmark key to accept. 55 00:02:48,821 --> 00:02:51,208 Press the main key to set the frequency. 56 00:02:51,210 --> 00:02:53,488 In this case we know the correct frequency 57 00:02:53,490 --> 00:02:59,080 and we will set it using the knob. 58 00:02:59,213 --> 00:03:01,875 The NORTEC 600 includes a useful slide rule tool 59 00:03:02,008 --> 00:03:04,271 for determining the standard depth of penetration 60 00:03:04,273 --> 00:03:07,356 for a given material at a specified frequency. 61 00:03:07,358 --> 00:03:09,600 The user can select a material from the list 62 00:03:09,601 --> 00:03:11,235 and modify the frequency 63 00:03:11,236 --> 00:03:13,675 to determine the depth of penetration. 64 00:03:13,808 --> 00:03:15,303 To access the slide rule menu, 65 00:03:15,438 --> 00:03:18,736 press the MAIN key until slide rule appears at the E key. 66 00:03:18,871 --> 00:03:22,993 Press the E key and the slide rule menu is displayed. 67 00:03:23,126 --> 00:03:26,891 Place the calibration standard with the undamaged fastener to the left 68 00:03:26,893 --> 00:03:29,220 and the damaged fastener to the right. 69 00:03:29,221 --> 00:03:31,216 Place the probe over the good fastener 70 00:03:31,218 --> 00:03:33,425 and press and hold the null key. 71 00:03:33,426 --> 00:03:34,918 When the Nortec beeps, 72 00:03:34,920 --> 00:03:36,920 lift the probe straight into the air. 73 00:03:36,921 --> 00:03:38,795 Avoid tilting the probe. 74 00:03:38,796 --> 00:03:42,878 Place the probe over the good fastener and press null. 75 00:03:42,880 --> 00:03:44,376 Lift the probe and place it 76 00:03:44,378 --> 00:03:47,148 over the damaged fastener and press freeze. 77 00:03:47,150 --> 00:03:48,871 Once the screen is frozen, 78 00:03:48,873 --> 00:03:50,758 you can change the angle. 79 00:03:50,760 --> 00:03:54,080 Try to set the liftoff as horizontally as possible. 80 00:03:54,081 --> 00:03:56,956 Use the knob to fine tune angle. 81 00:03:56,958 --> 00:03:58,696 To change from coarse to fine tuning, 82 00:03:58,698 --> 00:04:01,423 press the checkmark key. 83 00:04:01,425 --> 00:04:04,623 Press Gain three times to change the V gain. 84 00:04:04,625 --> 00:04:06,538 Set the gain so that the damaged signal 85 00:04:06,540 --> 00:04:10,961 reaches 90% of the screen height. 86 00:04:10,963 --> 00:04:15,491 Press Freeze and Full Next to enter full screen mode. 87 00:04:15,493 --> 00:04:17,416 Place the probe over the good fastener, 88 00:04:17,418 --> 00:04:19,330 and then the damaged fastener. 89 00:04:19,331 --> 00:04:22,255 Both fasteners should be separated by height only. 90 00:04:22,256 --> 00:04:24,485 If the damaged fastener is too far to the left, 91 00:04:24,486 --> 00:04:26,076 increase the frequency. 92 00:04:26,078 --> 00:04:28,128 If the damaged fastener is too far to the right, 93 00:04:28,130 --> 00:04:30,493 decrease the frequency. 94 00:04:30,495 --> 00:04:32,771 Changing D Erase and Persist settings 95 00:04:32,773 --> 00:04:35,036 can be useful for some applications. 96 00:04:35,038 --> 00:04:37,666 D Erase erases the eddy current display 97 00:04:37,668 --> 00:04:39,868 at user-determined time intervals. 98 00:04:39,870 --> 00:04:42,943 Persist erases all information at a particular point 99 00:04:42,945 --> 00:04:45,970 after the selected amount of time has expired. 100 00:04:45,971 --> 00:04:48,916 To set the D Erase and Persist time intervals 101 00:04:48,918 --> 00:04:50,715 press DISP twice, 102 00:04:50,716 --> 00:04:52,773 then use the A and B keys and knob 103 00:04:52,775 --> 00:04:55,816 to adjust the values. 104 00:04:55,818 --> 00:04:59,910 Alternate Method to inspect Subsurface Cracks 105 00:04:59,911 --> 00:05:01,971 Another way to inspect sub-surface cracks 106 00:05:01,973 --> 00:05:03,948 is to use a sliding probe. 107 00:05:03,950 --> 00:05:07,030 Set your frequency to 2.5kHz. 108 00:05:07,031 --> 00:05:11,396 Place the sliding probe on a defect free area of the calibration standard. 109 00:05:11,398 --> 00:05:13,416 Press and hold the null key. 110 00:05:13,418 --> 00:05:16,175 This will activate the auto lift off function. 111 00:05:16,176 --> 00:05:19,043 When you hear the beep, lift the probe. 112 00:05:19,045 --> 00:05:22,891 Place the probe back on the defect free area and press null. 113 00:05:22,893 --> 00:05:25,900 Slikde the probe to scan an undamaged rivet. 114 00:05:25,901 --> 00:05:27,506 Press freeze. 115 00:05:27,508 --> 00:05:29,340 Using the angle key and knob, 116 00:05:29,341 --> 00:05:31,841 Adjust the signal so that the starting and end points 117 00:05:31,843 --> 00:05:38,940 are both parallel. 118 00:05:38,941 --> 00:05:40,806 Press gain twice and use the knob 119 00:05:40,808 --> 00:05:43,013 to adjust the horizontal gain. 120 00:05:43,015 --> 00:05:46,608 The H gain signal should cover 5 divisions. 121 00:05:46,610 --> 00:05:50,320 Press gain again and use the knob to adjust the vertical gain. 122 00:05:50,321 --> 00:05:53,563 The V gain signal should cover 3 divisions. 123 00:05:53,565 --> 00:05:55,705 Press freeze and full next. 124 00:05:55,706 --> 00:05:59,506 Place the probe on a defect free area and press null. 125 00:05:59,508 --> 00:06:01,405 Scan the standard. 126 00:06:01,406 --> 00:06:03,695 If desired, adjust the alarm box 127 00:06:03,696 --> 00:06:05,158 for better separation between 128 00:06:05,160 --> 00:06:08,935 damaged and undamaged fastener signals. 129 00:06:08,936 --> 00:06:11,006 We hope you enjoyed this instructional video 130 00:06:11,008 --> 00:06:12,931 on how to configure the NORTEC 600 131 00:06:12,933 --> 00:06:15,675 to inspect surface and sub-surface cracks. 132 00:06:15,676 --> 00:06:18,545 For more information on Olympus eddy current solutions, 133 00:06:18,546 --> 00:06:20,555 contact your local representative 134 00:06:20,556 --> 00:06:28,043 or visit us online at www.olympus-ims.com.