Inspection & Measurement Systems

应用

Application Notes

Ultrasonic Preamplifiers

Application: Use of preamplifiers in ultrasonic flaw detection, thickness gaging, and other ultrasonic test and measurement applications.

Background: Certain ultrasonic applications require the addition of a separate preamplifier to standard flaw detectors, thickness gages, or other instruments in order to provide sufficient gain or broadband signal-to-noise enhancement for optimum echo acquisition. This sometimes happens in industrial applications that involve testing thick sections of material that exhibit high ultrasonic attenuation or scattering, such as nodular cast iron, austenitic steel, brass, bronze, rubber, fiberglass, or certain composites. A preamplifier is also an important component of acoustic emission test systems that require amplification of low amplitude signals. Preamps can also be used to compensate for amplitude losses caused by very long cables between the transducer and the test instrument, to improve signal-to-noise ratio in scanning applications involving large test specimens, and to amplify small tip diffracted signals in time-of-flight diffraction (TOFD) tests. Panametrics-NDT preamplifiers have also been used in many diverse research applications such as elastic modulus measurements in rocks and minerals, analysis of acoustic scatterers such as bubbles or plankton in water and other liquids, and biological tissue analysis. In general, a preamplifier can be used in almost any application where additional gain is required for optimum performance.

Equipment: There are seven models of Panametrics-NDT preamplifier, which offer bandwidth and gain choices as listed below. All of these models are compatible with a wide variety of flaw detectors, thickness gages, and pulser/receivers. For full specifications, please refer to the Preamplifier Data Sheet.

Model

5660B

5660C

5662

5670

5676

5678

5682

Bandwidth

20 kHz to
2 MHz

500 Hz to
2 MHz

50 kHz to
5 MHz

50 kHz to
10 MHz

50 kHz to
20 MHz

200 kHz to
40 MHz

500 kHz to
25 MHz

Gain

40/60 dB

40/60 dB

34/54 dB

40 dB

40 dB

40 dB

30 dB


Theory of Operation: The primary purpose of a preamplifier is to improve receiver signal to noise ratios in applications involving low level acoustic signals. By providing supplementary low noise amplification, a preamp allows use of lower gain in the primary test instrument, thus limiting baseline noise. A preamp may permit detection of small echoes that cannot be detected even at maximum gain in the primary test instrument. Additionally, the RF bandpass filtering in preamplifiers further helps limit high frequency noise.

The waveforms below represents a thickness measurement of a 5.6 in. (140 mm) rubber tire tread over fabric reinforcements in a heavy equipment tire, using a special 500 KHz dual element transducer. In the absence of a preamplifier in some challenging tests like this, instrument receiver gain must be set near maximum in order to obtain an echo, and there can be significant baseline noise:

Screen Image - no preamp

But if a preamplifier is added to the receiver side of the system, then instrument gain can be lowered by more than 30 dB while still maintaining a strong backwall echo, greatly reducing baseline noise:



Connection Procedure:
There are two ways of connecting a preamplifier into a test system, as outlined below.
1. Pulse/echo systems with single element transducers


Figure 1 -- Cable connections for pulse/echo setup.

When a preamplifier is used with a single element transducer in a pulse-echo system, the transducer cable should be run into a "T" connector. One branch of the T is connected to the transmitter transducer connector of the ultrasonic instrument. The other branch of the T goes to the preamplifier input. The preamplifier output should be connected to the receiver transducer connector of the ultrasonic instrument. If the ultrasonic instrument has a mode selection switch (single element/dual element or pulse/echo- through transmission), it should be set to the dual element or thru transmission position, even though a single element transducer is being used.

2. Through transmission, pitch/catch, or dual element transducer systems



Figure 2 -- Cable connections for through transmission system.

For systems using separate transmitter and receiver transducers or dual-element probes, the preamplifier is connected on the receiver side of the system between the receiver transducer and the receiver input of the test instrument. If the ultrasonic instrument has a mode selection switch (single element/dual element or pulse/echo-thru transmission), it should be set to the dual element or thru transmission position.

General notes: As noted above, whenever using preamplifiers the test instrument should normally be set to through transmission, dual element, or pitch/catch mode even when using single element transducers in a pulse/echo setup.

Preamplifiers are most effective in situations where signal loss is due to attenuation rather than scattering from grain boundaries, reinforcing fibers, or other discrete reflectors. This is because scatter noise will be amplified along with the desired echoes.

The amount of gain that can be effectively used in any given ultrasonic system is limited by the internal noise levels and amplifier characteristics of the instruments involved. Panametrics-NDT preamplifiers have very low internal noise (5µv peak to peak referred to the input for the 5660B), but generally system RF voltage gains above 100 dB are not practical. Panametrics-NDT flaw detectors and pulser/receivers should be used in the lowest practical gain setting when a preamplifier is used to avoid amplifier saturation.

In some cases, a high gain preamplifier output can overload input circuitry of certain ultrasonic instruments, resulting in baseline offset and non-linear amplitude presentations. Careful adjustment of RF filtering and instrument gain can help minimize this effect. In extreme cases an external attenuator can be inserted between the preamplifier output and the instrument receiver input.

In high gain systems, it is important to insure that the coaxial integrity of the transducer cables is maintained in order to prevent external RF noise pickup.

When operating in pulse/echo mode, the transducer excitation pulse must not exceed 500 volts for the Model 5682, or 300 volts for all other models, to prevent possible preamplifier damage. This restriction does not apply to through transmission or dual element setups.

When operating in pulse/echo mode, amplification of the excitation pulse or interface echoes can result in reduced near surface resolution. Careful adjustment of RF filtering and pulser damping can help minimize this effect.

When using a preamplifier to compensate for transmission losses in very long cables, the preamp must be located at the transducer end of the cable.

Продукты, используемые для этой цели

5058PR

5058PR高压超声脉冲发生器接收器专用于对声束衰减性强的材料的检测和测量应用。其脉冲发生器最高可发射900伏激励脉冲,从而可适当对低频探头进行补偿。

5072PR

5072PR手动控制超声脉冲发生器接收器既可用于常规应用,又可用于高频应用。用于一般检测的理想配置是35 MHz(-3 dB)带宽和尖脉冲发生器。 优化的宽带激励会产生极佳的时域恢复效果,特别是在15~30 MHz的情况下。

5073PR

5073PR手动控制超声脉冲发生器接收器既可用于常规应用,又可用于高频应用。 在对轴向和近表面分辨率要求极高的应用中,这款设备的75 MHz(-3 dB)带宽和快速上升的尖脉冲发生器可提高50 MHz探头的工作性能。

5077PR

5077PR手动控制超声脉冲发生器接收器具有35 MHz(-3 dB)带宽和方波脉冲发生器接收器,是一款可使散射材料的响应最大化的理想设备。方波脉冲发生器在与10 MHz或10 MHz以下的探头一起使用进行检测时,会发挥出其明显的优势。

EPOCH 1000系列

EPOCH 1000是一款高级常规超声探伤仪,可在授权的Olympus服务中心为其升级,使其具有相控阵成像功能。其主要特性包括:符合EN12668-1标准、37个数字式接收器滤波器选择、用于高速扫查的6 kHz脉冲重复频率。

EPOCH 600NEW

EPOCH 600是一款中级水平的手持式超声探伤仪。EPOCH 600仅重1.68公斤(3.72磅),机身横宽竖窄,机壳坚固耐用,可以在非常恶劣的检测环境中正常工作。EN12668-1型号仪器增加了400 V PerfectSquare可调方波脉冲发生器及用于加强信噪比的数字式滤波性能。

EPOCH LT

EPOCH LT是一款手持式初级数字式探伤仪,厚度仅为38毫米(1.5英寸),重量仅为1公斤(2.35磅),机身结构横窄竖宽,具有最小60 Hz的快速显示更新率、自动探头校准、数字式数据记录器等特点,以及许多其他测量功能。

EPOCH LTC

EPOCH LTC是一款中级水平的手持式超声探伤仪。这款小巧的仪器为横窄竖宽结构,仅重0.96公斤(2.12磅)。这款功能齐全的仪器符合EN12668-1标准,不仅具有各种标准的功能,还可附加满足用户的具体检测需求的特殊选项。

EPOCH XT

EPOCH XT是一款高级便携式超声探伤仪。具有很多标准测量功能,其中包括可调方波脉冲发生器、可选窄带和宽带数字式滤波器、0~110 dB的增益范围、峰值记忆和峰值保持、可调PRF。

English | 日本語 | français | 简体中文 | Deutsch | italiano | čeština | magyar | Tiểng Việt | Español | русский | polski | português | 한국어