Remote Visual Inspection

Specifications

SCOPE VARIATION
Model No.Scope
Diameter
Scope LengthTube FlexibilityArticulation Angle
UP/DOWN/
RIGHT/LEFT
Approx. System Weight
(with battery)
IPLEX LXIPLEX LT
IV8420L1 IV8420L2 Ø4.0 mm
(0.157 in)
2.0 m
(6.6 ft)
Uniformed stiffness 130° 2.7 kg
(6.0 lb)
IV8435L1 IV8435L2 3.5 m
(11.5 ft)
2.8 kg
(6.2 lb)
IV8620L1 IV8620L2 Ø6.0 mm (0.236 in) 2.0 m
(6.6 ft)
Tough Tapered Flex™ tube with flexibility gradually increasing toward the distal end. 150°
IV8635L1 IV8635L2 3.5 m
(11.5 ft)
130° 2.9 kg
(6.4 lb)
IV8675L1 IV8675L2 7.5 m
(24.6 ft)
90° 3.2 kg
(7.1 lb)
IV8835L1 IV8835L2 Ø8.5 mm (0.335 in) 3.5 m
(11.5 ft)
140° 3.0 kg
(6.6 lb)
IV88100L1 IV88100L2 10.0 m
(32.8 ft)
115° 3.9 kg
(9.0 lb)
IV88200L1 IV88200L2 20.0 m
(65.6 ft)
Uniformed stiffness 70° 5.2 kg
(11.5 lb)

IPLEX LX IPLEX LT
INSERTION TUBE
Articulation Mechanics TrueFeel™ scope tip articulation with electronic power-assisted, manual articulation
Temperature Sensor 2-stage indicator for high temperature warning N/A
Illumination Multiple ultra-high intensity LEDs,
2-stage setting for extra-bright illumination
Multiple ultra-high intensity LEDs
Optical Tip Adapter Identification1 SmartTip™ Automatic Recognition Technology N/A
Optical Tip Adapter Locking Mechanism Double Threaded Attachment, with O-ring seal
HANDHELD CONTROL UNIT
Dimensions (W x D x H) 311 x 92 x 192mm
Weight Approx. 750g
Quick access buttons: The following features have dedicated quick access buttons, levers or joysticks on the hand control unit: Live mode, WiDER"! Gain Control (IPLEX LX only), Brightness, Zoom, Freeze Image, Record Still/Video, Articulation Control, Articulation Lock, Menu Access, Saved Recording Index Access, Quick Last Image Recall.
MAIN UNIT
Dimensions (W x D x H) 227 x 64 x 189 mm
Maximum Dimensions: 239 x 99 x 215 mm
LCD monitor 6.5 in. daylight-view LCD, anti-reflective type
Power Supply Battery: 10.8 V nominal
IV84 and IV86 Series: over 120-minute operating time, IV88 Series: over 100-minute operating time
AC power: 100 V to 240 V, 50/60 Hz (with supplied AC adapter)
Output Connector RCA composite video output
USB Connector Type A connector, Version 2.0 standards
SOFTWARE MANIPULATIONS
Image Manipulations 3 step digital zoom, 8 step digital brightness control
Adjustable Gain Control 3 step adjustable gain control with exclusive WiDER™ brightness level balancing technology N/A
Display Text Options 30-character title display
RECORDING MANAGEMENT FUNCTIONS
Recording Media USB flash drive (1GB standard, compatible with recommended USB flash device)
Overlay Selectable 30 character title with date, time and system setting
Thumbnail image display Recorded images can be displayed as thumbnails
Still image Recording Resolution - NTSC: H640 x V480 (Pixel), PAL: H768 x V576 (Pixel)
Recording Format - Compressed JPEG format
Size of a Single Image - Approx. 300 KB (Approx. 3,400 images per 1 GB memory)
Video Recording Resolution - H640 x V480 (Pixel)
Recording Format - AVI M-JPEG format, Can be played on the Windows Media Player.
Size of a Single Image - Approx. 500 KB per second (Approx. 30 min. per 1 GB memory)
STEREO MEASUREMENT FUNCTIONS - Optional function for IPL EX LX
Distance Distance between two points Note: Stereo Measurement functions are available by upgrading the IPLEX LX.
Point-to-Line Perpendicular distance between a point and a user-defined line
Depth Orthogoral depth/height distance between a point and a user-defined plane

1. On the 8.5 mm scope tip, the SmartTip function is limited to recognition of the tip adapter attachment to the scope only and  manual selection of the tip adapter model is required by the user.

OPERATING ENVIRONMENT
Operating Temperature Insertion Tube In air: -25 ºC to 100 ºC IPLEX LX IV84 and IV86 series
-25 ºC to 80 ºC IPLEX LT and IPLEX LX IV88 series
In water: 10 ºC to 30 ºC (50 ºF to 86 ºF)
Other parts than above In air: -10 °C to 40 ºC (14 ºF to 104 ºF)
Operating Atmospheric Pressure Insertion Tube In air: 1013 hPa
In water: IV84 series - 1013 to 1368 hPa
IV86 series - 1013 to 1772 hPa
IV88 series - 1013 to 2026 hPa
Other parts than above In air: 1013 hPa
Relative Humidity All parts 15 % to 90 %
Liquid Resistance All parts Operable when exposed to machine oil, light oil or 5 % saline solution.
Waterproofing Insertion Tube Operable under water with viewing tip adapter attached.
Not operable underwater with stereo measurement tip adapters.
Other parts than above Operable in blowing rain conditions (Battery compartment must be closed).
Not operable underwater.
MIL-STD COMPLIANCE
The operating environment performance is confirmed by the following MIL-STD-810G and MIL-STD-461F under battery operation.
TypeMethod
Vibration1 MIL-STD-810G, Method 514.5, Procedure I (General vibration test).
Shock MIL-STD-810G, Method 516.5, Procedure IV (Transit drop test).
Water Resistance1 MIL-STD-810G, Method 506, Procedure I (Rain and Blowing rain test).
Humidity1 MIL-STD-810G, Method 507.4
Salt Fog1 MIL-STD-810G, Method 509
Sand and Dust1 MIL-STD-810G, Method 510.4, Procedure I (Blowing dust test).
Icing/Freezing Rain1 MIL-STD-810G, Method 521.2
Electromagnetic Interference (EMI)2 MIL-STD-461F, RS-103, (Radiated susceptibility test for non-metallic above deck shipboard conditions).
Explosive Atmosphere1 MIL-STD-810G, Method 511.4, Procedure I (Operation in an explosive atmosphere test).

*1. Excluding IV88 series.
*2. Excluding IV8675L1, IV8675L2 and IV88 series.

OPTICAL TIP adapter VARIATION
Optical AdapterOptical systemDistal end
Field of viewDirection
of view
Depth of field1Outer diameter2Distal end3
6.0 mm Viewing Tip adapters AT40D-IV86 40º Forward 200 to ∞ mm ø6.0 mm 19.8 mm
AT80D/NF-IV86 80º 8 to ∞ mm
AT80D/FF-IV86 80º 35 to ∞ mm
AT120D/NF-IV86 120º 4 to 190 mm
AT120D/FF-IV86 120º 25 to ∞ mm
AT80S-IV86 80º Side 18 to ∞ mm 25.2 mm
AT120S/NF-IV86 120º 1 to 25 mm
AT120S/FF-IV86 120º 5 to ∞ mm
4.0 mm Viewing Tip Adapters AT120D/NF-IV84 120º Forward 4 to 190 mm ø4.0 mm 19.7 mm
AT120D/FF-IV84 120º 25 to ∞ mm 19.6 mm
AT120S/NF-IV84 120º Side 1 to 20 mm 22.2 mm
AT120S/FF-IV84 120º 6 to ∞ mm
8.5 mm Viewing Tip Adapters AT80D/FF-IV88 80º Forward 35 to ∞ mm ø8.5 mm 22.9 mm
AT120D/NF-IV88 120º 4 to 190 mm
AT120D/FF-IV88 120º 25 to ∞ mm 29.5 mm
AT120S/NF-IV88 120º Side 1 to 25 mm
AT120S/FF-IV88 120º 5 to ∞ mm
Stereo Tip Adapters
(4.0 mm, 6.0 mm, and 8.5 mm)
AT50D/50D-IV84 50º/50º Forward/Forward 5 to ∞ mm ø4.0 mm 25.0 mm
AT50S/50S-IV84 50º/50º Side/Side 4 to ∞ mm 28.8 mm
AT60D/60D-IV86 60º/60º Forward/Forward 5 to ∞ mm ø6.0 mm 25.9 mm
AT60S/60S-IV86 60º/60º Side/Side 4 to ∞ mm 32.2 mm
AT60D/60D-IV88 60º/60º Forward/Forward 5 to ∞ mm ø8.5 mm 28.9 mm
AT60S/60S-IV88 60º/60º Side/Side 4 to ∞ mm 36.5 mm

1. Indicates the viewing distance with optimal focus.
2. The adapter can be inserted into a o4.0 mm, o6.0 mm or o8.5 mm hole when it is mounted on the scope.
3. Indicates the length of the rigid portion at the scope's distal end when mounted.


Inspection & Measurement Systems
English | 日本語 | français | 简体中文 | Deutsch | italiano | čeština | magyar | Tiểng Việt | Español | русский | polski | português | 한국어

Advanced NDT Solutions

Flaw Detectors

Integrated Inspection Systems