| SCOPE VARIATION | ||||||
| Model No. | Scope
Diameter | Scope Length | Tube Flexibility | Articulation Angle
UP/DOWN/ RIGHT/LEFT | Approx. System Weight
(with battery) | |
| IPLEX LX | IPLEX LT | |||||
| IV8420L1 | IV8420L2 |
Ø4.0 mm
(0.157 in) |
2.0 m
(6.6 ft) | Uniformed stiffness | 130° |
2.7 kg
(6.0 lb) |
| IV8435L1 | IV8435L2 |
3.5 m
(11.5 ft) |
2.8 kg
(6.2 lb) | |||
| IV8620L1 | IV8620L2 | Ø6.0 mm (0.236 in) |
2.0 m
(6.6 ft) | Tough Tapered Flex™ tube with flexibility gradually increasing toward the distal end. | 150° | |
| IV8635L1 | IV8635L2 |
3.5 m
(11.5 ft) | 130° |
2.9 kg
(6.4 lb) | ||
| IV8675L1 | IV8675L2 |
7.5 m
(24.6 ft) | 90° |
3.2 kg
(7.1 lb) | ||
| IV8835L1 | IV8835L2 | Ø8.5 mm (0.335 in) |
3.5 m
(11.5 ft) | 140° |
3.0 kg
(6.6 lb) | |
| IV88100L1 | IV88100L2 |
10.0 m
(32.8 ft) | 115° |
3.9 kg
(9.0 lb) | ||
| IV88200L1 | IV88200L2 |
20.0 m
(65.6 ft) | Uniformed stiffness | 70° |
5.2 kg
(11.5 lb) | |
| IPLEX LX | IPLEX LT | |
| INSERTION TUBE | ||
| Articulation Mechanics | TrueFeel™ scope tip articulation with electronic power-assisted, manual articulation | |
| Temperature Sensor | 2-stage indicator for high temperature warning | N/A |
| Illumination |
Multiple ultra-high intensity LEDs,
2-stage setting for extra-bright illumination | Multiple ultra-high intensity LEDs |
| Optical Tip Adapter Identification1 | SmartTip™ Automatic Recognition Technology | N/A |
| Optical Tip Adapter Locking Mechanism | Double Threaded Attachment, with O-ring seal | |
| HANDHELD CONTROL UNIT | ||
| Dimensions (W x D x H) | 311 x 92 x 192mm | |
| Weight | Approx. 750g | |
| Quick access buttons: | The following features have dedicated quick access buttons, levers or joysticks on the hand control unit: Live mode, WiDER"! Gain Control (IPLEX LX only), Brightness, Zoom, Freeze Image, Record Still/Video, Articulation Control, Articulation Lock, Menu Access, Saved Recording Index Access, Quick Last Image Recall. | |
| MAIN UNIT | ||
| Dimensions (W x D x H) |
227 x 64 x 189 mm
Maximum Dimensions: 239 x 99 x 215 mm | |
| LCD monitor | 6.5 in. daylight-view LCD, anti-reflective type | |
| Power Supply |
Battery: 10.8 V nominal
IV84 and IV86 Series: over 120-minute operating time, IV88 Series: over 100-minute operating time AC power: 100 V to 240 V, 50/60 Hz (with supplied AC adapter) | |
| Output Connector | RCA composite video output | |
| USB Connector | Type A connector, Version 2.0 standards | |
| SOFTWARE MANIPULATIONS | ||
| Image Manipulations | 3 step digital zoom, 8 step digital brightness control | |
| Adjustable Gain Control | 3 step adjustable gain control with exclusive WiDER™ brightness level balancing technology | N/A |
| Display Text Options | 30-character title display | |
| RECORDING MANAGEMENT FUNCTIONS | ||
| Recording Media | USB flash drive (1GB standard, compatible with recommended USB flash device) | |
| Overlay | Selectable 30 character title with date, time and system setting | |
| Thumbnail image display | Recorded images can be displayed as thumbnails | |
| Still image Recording | Resolution - NTSC: H640 x V480 (Pixel), PAL: H768 x V576 (Pixel) | |
| Recording Format - Compressed JPEG format | ||
| Size of a Single Image - Approx. 300 KB (Approx. 3,400 images per 1 GB memory) | ||
| Video Recording | Resolution - H640 x V480 (Pixel) | |
| Recording Format - AVI M-JPEG format, Can be played on the Windows Media Player. | ||
| Size of a Single Image - Approx. 500 KB per second (Approx. 30 min. per 1 GB memory) | ||
| STEREO MEASUREMENT FUNCTIONS - Optional function for IPL EX LX | ||
| Distance | Distance between two points | Note: Stereo Measurement functions are available by upgrading the IPLEX LX. |
| Point-to-Line | Perpendicular distance between a point and a user-defined line | |
| Depth | Orthogoral depth/height distance between a point and a user-defined plane | |
1. On the 8.5 mm scope tip, the SmartTip function is limited to recognition of the tip adapter attachment to the scope only and manual selection of the tip adapter model is required by the user.
| OPERATING ENVIRONMENT | ||
| Operating Temperature | Insertion Tube |
In air: -25 ºC to 100 ºC IPLEX LX IV84 and IV86 series
-25 ºC to 80 ºC IPLEX LT and IPLEX LX IV88 series In water: 10 ºC to 30 ºC (50 ºF to 86 ºF) |
| Other parts than above | In air: -10 °C to 40 ºC (14 ºF to 104 ºF) | |
| Operating Atmospheric Pressure | Insertion Tube |
In air: 1013 hPa
In water: IV84 series - 1013 to 1368 hPa IV86 series - 1013 to 1772 hPa IV88 series - 1013 to 2026 hPa |
| Other parts than above | In air: 1013 hPa | |
| Relative Humidity | All parts | 15 % to 90 % |
| Liquid Resistance | All parts | Operable when exposed to machine oil, light oil or 5 % saline solution. |
| Waterproofing | Insertion Tube |
Operable under water with viewing tip adapter attached.
Not operable underwater with stereo measurement tip adapters. |
| Other parts than above |
Operable in blowing rain conditions (Battery compartment must be closed).
Not operable underwater. | |
| MIL-STD COMPLIANCE | |
| The operating environment performance is confirmed by the following MIL-STD-810G and MIL-STD-461F under battery operation. | |
| Type | Method |
| Vibration1 | MIL-STD-810G, Method 514.5, Procedure I (General vibration test). |
| Shock | MIL-STD-810G, Method 516.5, Procedure IV (Transit drop test). |
| Water Resistance1 | MIL-STD-810G, Method 506, Procedure I (Rain and Blowing rain test). |
| Humidity1 | MIL-STD-810G, Method 507.4 |
| Salt Fog1 | MIL-STD-810G, Method 509 |
| Sand and Dust1 | MIL-STD-810G, Method 510.4, Procedure I (Blowing dust test). |
| Icing/Freezing Rain1 | MIL-STD-810G, Method 521.2 |
| Electromagnetic Interference (EMI)2 | MIL-STD-461F, RS-103, (Radiated susceptibility test for non-metallic above deck shipboard conditions). |
| Explosive Atmosphere1 | MIL-STD-810G, Method 511.4, Procedure I (Operation in an explosive atmosphere test). |
*1. Excluding IV88 series.
*2. Excluding IV8675L1, IV8675L2 and IV88 series.
| OPTICAL TIP adapter VARIATION | ||||||
| Optical Adapter | Optical system | Distal end | ||||
| Field of view | Direction
of view | Depth of field1 | Outer diameter2 | Distal end3 | ||
| 6.0 mm Viewing Tip adapters | AT40D-IV86 | 40º | Forward | 200 to ∞ mm | ø6.0 mm | 19.8 mm |
| AT80D/NF-IV86 | 80º | 8 to ∞ mm | ||||
| AT80D/FF-IV86 | 80º | 35 to ∞ mm | ||||
| AT120D/NF-IV86 | 120º | 4 to 190 mm | ||||
| AT120D/FF-IV86 | 120º | 25 to ∞ mm | ||||
| AT80S-IV86 | 80º | Side | 18 to ∞ mm | 25.2 mm | ||
| AT120S/NF-IV86 | 120º | 1 to 25 mm | ||||
| AT120S/FF-IV86 | 120º | 5 to ∞ mm | ||||
| 4.0 mm Viewing Tip Adapters | AT120D/NF-IV84 | 120º | Forward | 4 to 190 mm | ø4.0 mm | 19.7 mm |
| AT120D/FF-IV84 | 120º | 25 to ∞ mm | 19.6 mm | |||
| AT120S/NF-IV84 | 120º | Side | 1 to 20 mm | 22.2 mm | ||
| AT120S/FF-IV84 | 120º | 6 to ∞ mm | ||||
| 8.5 mm Viewing Tip Adapters | AT80D/FF-IV88 | 80º | Forward | 35 to ∞ mm | ø8.5 mm | 22.9 mm |
| AT120D/NF-IV88 | 120º | 4 to 190 mm | ||||
| AT120D/FF-IV88 | 120º | 25 to ∞ mm | 29.5 mm | |||
| AT120S/NF-IV88 | 120º | Side | 1 to 25 mm | |||
| AT120S/FF-IV88 | 120º | 5 to ∞ mm | ||||
|
Stereo Tip Adapters
(4.0 mm, 6.0 mm, and 8.5 mm) | AT50D/50D-IV84 | 50º/50º | Forward/Forward | 5 to ∞ mm | ø4.0 mm | 25.0 mm |
| AT50S/50S-IV84 | 50º/50º | Side/Side | 4 to ∞ mm | 28.8 mm | ||
| AT60D/60D-IV86 | 60º/60º | Forward/Forward | 5 to ∞ mm | ø6.0 mm | 25.9 mm | |
| AT60S/60S-IV86 | 60º/60º | Side/Side | 4 to ∞ mm | 32.2 mm | ||
| AT60D/60D-IV88 | 60º/60º | Forward/Forward | 5 to ∞ mm | ø8.5 mm | 28.9 mm | |
| AT60S/60S-IV88 | 60º/60º | Side/Side | 4 to ∞ mm | 36.5 mm | ||
1. Indicates the viewing distance with optimal focus.
2. The adapter can be inserted into a o4.0 mm, o6.0 mm or o8.5 mm hole when it is mounted on the scope.
3. Indicates the length of the rigid portion at the scope's distal end when mounted.
Copyright 2011 OLYMPUS CORPORATION, All rights reserved. Terms Of Use | Privacy Statement