| Measurement Mode: Pulse echo with dual element transducers
|
ThicknessMeasurement Range: 0.020 to 19.999" (0.50 to 508 mm)
Note: Thickness range depends on material, transducer type, surface condition, and temperature.
|
MaterialVelocity Calibration Range: 0.0300-0.5511 in/µsec
(0.762- 13.999 mm/µsec)
Thickness Display Resolution:
LOW: 0.01" 0.1 mm
STANDARD: 0.001" 0.01 mm
|
| Measurement Rates: Standard Rate: 4 per second
|
| Fast Rate: 20 per second
|
| Fast Min Mode: Rapid capture and recall of minimum thickness at Fast Rate of 20 per second.
|
| Freeze Mode: Freezes display to instantly capture critical thickness. Minimizes transducer couplant lift-off error and facilitates high temperature measurements.
|
| Automatic Probe Recognition: Automatically recognizes the listed Panametrics-NDT transducer types. Adjusts internal parameters and corrects V-path error.
|
| Zero Compensation Mode: Compensates for transducer temperature and zero offset.
|
| Display: 4 1/2 digit (19999 counts) Liquid Crystal Display (LCD), 0.4" (10.2 mm) numerals
|
| Display Hold/Blank Mode: Display holds or blanks after measurement.
|
| Electroluminescent Display Back Lighting: Selectable as "On" or "Auto On"
|
| Receiver Bandwidth: 1-15 MHz (-3dB)
|
| Metric/English Mode: Metric or English units
|
| Battery: 2 AA Alkaline Batteries
|
| Operating Time: 250 hours typical battery life; 30 hours continuous with backlight on
|
| Low Battery Indicator: Continuously indicates battery status
|
| BatterySaver: Auto Power Off/Continuous On
|
| Case/Keypad: Splash proof, impact-resistant Lexan® case. Sealed, color-coded keypad with tactile and audible feedback.
|
| Operating Temperature Range:-10° C to +50° C, 14° F to 122° F
|
| Size: 2.55" W x 5.05" L x 1.14" H(64.8 mm x 128.7 mm x 29 mm)
|
| Weight: 8.5 oz. (0.24 kg)
|
| Hazardous Area Operations: As defined by MIL-STD-810E, method 511.3, procedure 1.
|
|
*All specifications subject to change without notice.
|