The A-scan display is bright with great contrast and visibility. Every thickness reading in a stored B-scan can be reviewed on the 37DL PLUS and in the Interface Program. The user can select to view (Min/Max, Alarm, or A-scan flags) flags with Grid points. The file-based alphanumeric datalogger can use longer file names (32 character max) and ID numbers (20 character max). Grid files can be expanded by adding rows or columns or by changing the incrementing direction.
Transducers
The 37DL PLUS can be used with a vast selection of dual element and single element transducers. Its full line of D790 series dual element transducers feature Automatic Probe Recognition to provide maximum gage performance for each transducer. The 37DL PLUS also has the ability to use the E110-SB EMAT transducer for no-couplant steel thickness measurements through oxide-scaled surfaces. The Panametrics M2017 and M2091 transducers are ideally suited for measuring internal
oxide/scale build-up on boiler tubes. The 37DL PLUS is also compatible with the complete line of Panametrics Microscan single element direct contact, delay line, and immersion transducers ranging in frequency from 2 to 30 MHz. As a result, the gage can be used for non-corrosion applications involving materials such as plastic, fiberglass, composites, rubber, castings, rubber and glass. Application Auto-Recall automatically recalls 16 default and 10 custom Microscan transducer
setups from the gage's memory.
 |
Thru-Coat™ technology
This innovative feature uses a single backwall echo to measure true metal thickness. The Panametrics 37DL PLUS can display both metal and coating thickness, each adjusted for their correct material sound velocity. The gage can be set to display only the true metal thickness. There is no need to remove the paint or coating from the surface. Thru-Coat measurements use the new D7906-SM and D7908 dual element transducers
|
 |
Temperature Compensation
Variations in material temperature cause changes in sound velocity that may affect the accuracy of the thickness measurement. The Temperature Compensation feature allows the user to enter the calibration block temperature and manually input the current (high) temperature at the measurement points. The Panametrics 37DL PLUS will display the temperature corrected thickness that can then be saved to the internal datalogger.
|
 |
DB Grid View
The improved Grid View feature allows the user to select either a Min/Max, Alarm, or
A-scan flag to be viewed along with each grid point. This allows the user to quickly scan through a file and locate Min/Max, Alarm, or A-scan locations.
|
 |
Optional Oxide/Scale Measurement
This option uses advanced algorithms to measure the thickness of oxide/scale build-up on the inside of boiler tubes. The gage simultaneously displays the metal thickness of the boiler tube and the thickness of the oxide layer. Knowing the thickness of the oxide/scale helps predict tube life. We recommend the use of the M2017 or M2091 transducer for this application.
|
|
Measurements
Dual Element Measurement Mode:Time interval from a precision delay after the excitation pulse to the first echo.
Thru-Coat Measurement:Measurement of true metal and coating thickness
using a single backwall echo (with D7906-SM and D7908 transducers).
Thru-Paint Echo-to-Echo:Time
interval between two successive backwall echoes for elimination of paint or coating .
Single Element Measurement Mode:
Mode 1:Time interval between excitation pulse to first backwall echo.
Mode 2:Time interval between delay line echo to first backwall echo with delay or immersion transducers.
Mode 3:Time interval between successive backwall echoes following the first interface echo after the excitation pulse with delay line or immersion transducers.
Optional Oxide/Scale Measurement:
Measure metal thickness and internal oxide/scale build-up of pipes. Oxide/scale thickness range:
M2017: 0.25 mm to 1.25 mm (0.010 in. to 0.050 in.)
M2091: 0.15 mm to 1.25 mm (0.006 in. to 0.050 in.)
Thickness Range :
0.003 - 25.000 inches (0.080 - 635.00mm), depending on material, transducer surface condition, temperature, and SETUP selected
Material Velocity Range :
0.508 mm/µS to 13.998 mm/µS
0.020 inch/µS to 0.551 inch/µS
Resolution, selectable:
|
LOW:
|
0.1 mm
|
0.01 in.
| |
STANDARD:
|
0.01 mm
|
0.001 in.
|
Battery:
6V rechargeable NiCad battery pack, or field-replaceable alkaline AA batteries
BatteryLife:
25 hours in normal measurement mode with backlight off
Fast-Charger:
Two-hour Fast-Charger with universal voltage
TransducerFrequency Range:
2-30 MHz (-3 dB)
Operating Temperature:
-10° C to +50° C (+14° F to 122°F)
Keypad:
Sealed color-coded keypad with tactile and audible feedback
Case:
Impact-resistant and Water-resistant, gasketed Lexan® case with sealed connectors. Environmental IP-65 compliant
Liquid Crystal Display:
With backlight. Contrast keypad adjustable. Display area 102 mm x 86 mm (4.0 in. x 3.39 in.)
Hazardous Area Operations:
As defined by MIL-STD-810E, method 511.3, procedure 1
Rectification:
Full wave, RF, half wave positive, or half wave negative
DisplayRangeand Delay Control Metric/English Units
Size: (W x H x D)
138 mm x 238 mm x 38 mm (5.45 in. x 9.375 in. x 1.5 in.)
Weight:
0.95 Kg (2.1 lbs)
INTERNAL DATALOGGER
Datalogger and RS-232:
The 37DL PLUS identifies, stores, recalls, clears, and transmits thickness readings, waveform images, and gage setup information via the RS-232 serial port. Baud Rate, Word Length, Stop Bits, and Parity are adjustable from the keypad.
Maximum # of Stored Values:
Standard: 60,000 thickness readings
or 4,500 waveforms with thickness readings
File Names, ID# and Comments:
32 character file name plus 20-character alphanumeric location code with 4 comments per location
File Structures:
7 standard or custom application-specific file structures
Reports:
On-gage reporting of summary with statistics, Min/Max with locations, File Comparison, and Alarm Report. On-screen comparison of current and previous readings.
Standard Inclusions
Model 37DL PLUS Digital Ultrasonic Thickness Gage with Waveform Verification, AC or Battery Operation, 50-60 Hz
• D790-SM Transducer, 5.0 MHz, 0.434" (11mm) tip, temp. range -5° F to 932° F
• Charger/AC adapter (100, 115, 230 VAC)
• Internal Datalogger
• GageView Interface Program
• Test Block and Couplant
• RS-232 Cable
• Protective Pouch with Neck Strap
• Instruction Manual
• Two-Year Limited Warranty
• Measurement features: Thru-Coat, Echo-to-Echo Thru-Paint, EMAT Compatability, Min/Max Mode, Two Alarm Modes, Differential Mode, B-scan, Application Auto Recall, Temperature Compensation, Average/Min Mode
| Optional Accessories | | | 37DLP/EW |
Third-Year warranty
| | 36CA/EC-115 |
External Fast Charger, 115 VAC
| | 37DLP/SU/OXIDE |
Software option for the measurement of boiler tubes and internal oxide
| | 1/2XA/E110 |
Filter adapter for E110-SB EMAT transducer
|
*All specifications are subject to change without notice.
Some 37DL Plus Applications
We are sorry, we could not find an Application Note matching your search criteria. However, there is good chance we have a solution for your application. We suggest you contact your local sales representative who can advice you in selecting the best instrumentation for your application. You may also browse our product literature library. |