Inspection & Measurement Systems

Anwendungen

Anmerkungen zur Anwendung

Ultraschall-Vorverstärker

Application: Use of preamplifiers in ultrasonic flaw detection, thickness gaging, and other ultrasonic test and measurement applications.

Background: Certain ultrasonic applications require the addition of a separate preamplifier to standard flaw detectors, thickness gages, or other instruments in order to provide sufficient gain or broadband signal-to-noise enhancement for optimum echo acquisition. This sometimes happens in industrial applications that involve testing thick sections of material that exhibit high ultrasonic attenuation or scattering, such as nodular cast iron, austenitic steel, brass, bronze, rubber, fiberglass, or certain composites. A preamplifier is also an important component of acoustic emission test systems that require amplification of low amplitude signals. Preamps can also be used to compensate for amplitude losses caused by very long cables between the transducer and the test instrument, to improve signal-to-noise ratio in scanning applications involving large test specimens, and to amplify small tip diffracted signals in time-of-flight diffraction (TOFD) tests. Panametrics-NDT preamplifiers have also been used in many diverse research applications such as elastic modulus measurements in rocks and minerals, analysis of acoustic scatterers such as bubbles or plankton in water and other liquids, and biological tissue analysis. In general, a preamplifier can be used in almost any application where additional gain is required for optimum performance.

Equipment: There are seven models of Panametrics-NDT preamplifier, which offer bandwidth and gain choices as listed below. All of these models are compatible with a wide variety of flaw detectors, thickness gages, and pulser/receivers. For full specifications, please refer to the Preamplifier Data Sheet.

Model

5660B

5660C

5662

5670

5676

5678

5682

Bandwidth

20 kHz to
2 MHz

500 Hz to
2 MHz

50 kHz to
5 MHz

50 kHz to
10 MHz

50 kHz to
20 MHz

200 kHz to
40 MHz

500 kHz to
25 MHz

Gain

40/60 dB

40/60 dB

34/54 dB

40 dB

40 dB

40 dB

30 dB


Theory of Operation: The primary purpose of a preamplifier is to improve receiver signal to noise ratios in applications involving low level acoustic signals. By providing supplementary low noise amplification, a preamp allows use of lower gain in the primary test instrument, thus limiting baseline noise. A preamp may permit detection of small echoes that cannot be detected even at maximum gain in the primary test instrument. Additionally, the RF bandpass filtering in preamplifiers further helps limit high frequency noise.

The waveforms below represents a thickness measurement of a 5.6 in. (140 mm) rubber tire tread over fabric reinforcements in a heavy equipment tire, using a special 500 KHz dual element transducer. In the absence of a preamplifier in some challenging tests like this, instrument receiver gain must be set near maximum in order to obtain an echo, and there can be significant baseline noise:

Screen Image - no preamp

But if a preamplifier is added to the receiver side of the system, then instrument gain can be lowered by more than 30 dB while still maintaining a strong backwall echo, greatly reducing baseline noise:



Connection Procedure:
There are two ways of connecting a preamplifier into a test system, as outlined below.
1. Pulse/echo systems with single element transducers


Figure 1 -- Cable connections for pulse/echo setup.

When a preamplifier is used with a single element transducer in a pulse-echo system, the transducer cable should be run into a "T" connector. One branch of the T is connected to the transmitter transducer connector of the ultrasonic instrument. The other branch of the T goes to the preamplifier input. The preamplifier output should be connected to the receiver transducer connector of the ultrasonic instrument. If the ultrasonic instrument has a mode selection switch (single element/dual element or pulse/echo- through transmission), it should be set to the dual element or thru transmission position, even though a single element transducer is being used.

2. Through transmission, pitch/catch, or dual element transducer systems



Figure 2 -- Cable connections for through transmission system.

For systems using separate transmitter and receiver transducers or dual-element probes, the preamplifier is connected on the receiver side of the system between the receiver transducer and the receiver input of the test instrument. If the ultrasonic instrument has a mode selection switch (single element/dual element or pulse/echo-thru transmission), it should be set to the dual element or thru transmission position.

General notes: As noted above, whenever using preamplifiers the test instrument should normally be set to through transmission, dual element, or pitch/catch mode even when using single element transducers in a pulse/echo setup.

Preamplifiers are most effective in situations where signal loss is due to attenuation rather than scattering from grain boundaries, reinforcing fibers, or other discrete reflectors. This is because scatter noise will be amplified along with the desired echoes.

The amount of gain that can be effectively used in any given ultrasonic system is limited by the internal noise levels and amplifier characteristics of the instruments involved. Panametrics-NDT preamplifiers have very low internal noise (5µv peak to peak referred to the input for the 5660B), but generally system RF voltage gains above 100 dB are not practical. Panametrics-NDT flaw detectors and pulser/receivers should be used in the lowest practical gain setting when a preamplifier is used to avoid amplifier saturation.

In some cases, a high gain preamplifier output can overload input circuitry of certain ultrasonic instruments, resulting in baseline offset and non-linear amplitude presentations. Careful adjustment of RF filtering and instrument gain can help minimize this effect. In extreme cases an external attenuator can be inserted between the preamplifier output and the instrument receiver input.

In high gain systems, it is important to insure that the coaxial integrity of the transducer cables is maintained in order to prevent external RF noise pickup.

When operating in pulse/echo mode, the transducer excitation pulse must not exceed 500 volts for the Model 5682, or 300 volts for all other models, to prevent possible preamplifier damage. This restriction does not apply to through transmission or dual element setups.

When operating in pulse/echo mode, amplification of the excitation pulse or interface echoes can result in reduced near surface resolution. Careful adjustment of RF filtering and pulser damping can help minimize this effect.

When using a preamplifier to compensate for transmission losses in very long cables, the preamp must be located at the transducer end of the cable.

Products used for this application

5058PR

Der 5058PR Hochspannungs-Impulsgenerator-Empfänger ist speziell zum Prüfen und Messen von schallschwächenden Werkstoffen ausgelegt. Die Impulsgeneratoreinheit kann einem passenden niederfrequenten Schallkopf einen Anregungsimpuls bis 900 Volt liefern.

5072PR

Der handgesteuerte 5072PR Ultraschall-Impulsgenerator-Empfänger ist für normale und hochfrequente Anwendungen geeignet. Der Breitbandnadelimpuls von 35 MHz (-3 dB) ist ideal für allgemeine Prüfaufgaben. Optimale Breitbandanregung ergibt bessere Nahauflösung insbesondere zwischen 15-30 MHz.

5073PR

Der handgesteuerte 5073PR Ultraschall-Impulsgenerator-Empfänger ist für normale und hochfrequente Anwendungen geeignet. Die Bandbreite von 75 MHz (-3 dB) mit schnell ansteigendem Nadelimpuls erweitert die Leistung dieses Geräts auf Schallköpfe von 50 MHz in Anwendungsbereichen, die an der Grenze der Auflösung in Einschallrichtung und nahe der Oberfläche liegen.

5077PR

Der handgesteuerte 5077PR Ultraschall-Impulsgenerator-Empfänger ist für normale und hochfrequente Anwendungen geeignet. Mit einer Bandbreite von 35 MHz (-3 dB) und Rechteckimpuls ist dies Gerät ideal zur Optimierung der Leistung in schallstreuenden Werkstoffen. Der Rechteckimpuls ist besonders beim Prüfen mit Schallköpfen von 10 MHz und weniger von Vorteil.

Serie EPOCH 1000

Das EPOCH 1000 ist ein hochenwickeltes Prüfgerät für konventionellen Ultraschall, das in einem zugelassenen Olympus Kundendienststelle mit der Phased-Array-Darstellung aufgebessert werden kann. Seine wichtigesten Merkmale sind Konformität mit EN12668-1, 37 einstellbare Digitalfilter am Empfänger und Impulsfolgefrequenz von 6 kHz für schnelle Prüfung.

EPOCH 600NEW

Das EPOCH 600 ist ein Ultraschallhandprüfgerät der mittleren Klasse. Es wiegt nur 1,68 kg, sein horizontales Gehäuse widersteht sehr schwierigen Prüfumgebungen. Entspricht EN12668-1; weist unter anderem Funktionen wie den einstellbaren PerfectSquareTM Rechteckimpuls von 400 V und Digitalfilter auf, die das Signal-Rauschverhältnis verbessern.

EPOCH LT

Das EPOCH LT ist ein 38 mm dickes Handprüfgerät mit Grundfunktionen in einem nur 1 kg schweren vertikalen Gehäuse. Es besitzt schnelle Anzeigegeschwindigkeit mit mindestens 60 Hz, automatische Sensorjustierung, einen numerischen Messdatenspeicher und viele andere Messfunktionen.

EPOCH LTC

Das EPOCH LTC ist ein mittleres Ultraschallhandprüfgerät in einem kompakten, nur 0,96 kg schweren, vertikalen Gehäuse. Dies komplette Gerät, konform mit EN12668-1, mit einer breiten Auswahl an Standard- und speziellen Optionen entspricht Ihren Prüfanforderungen.

EPOCH XT

Das EPOCH XT ist ein hochentwickeltes tragbares Ultraschallprüfgerät mit vielen Standardfunktionen wie einstellbarem Rechteckimpuls, einstellbarem Schmalband- und Breitbanddigitalfilter, Verstärkungsbereich von 0 bis 110 dB, Spitzenwertspeicher und Halten des Spitzenwertes, einstellbarer IFF, 0,01 mm.

English | 日本語 | français | 简体中文 | Deutsch | italiano | čeština | magyar | Tiểng Việt | Español | русский | polski | português | 한국어