With federal and international regulations becoming increasingly stringent, the need to determine the sulfur content in the characterization of petroleum oils and fuels is a critical component. Energy Dispersive X-Ray Fluorescence, EDXRF, is a well established technique for the analysis
To meet these tough requirements for the measurement and control of sulfur in fuels and other petroleum products, Olympus NDT has developed the X-5000. The X-5000 is also ideal for measuring other elements, like Fe, Ni, V, Pb, Cr, Cu, and Cl in crude oil, as well as other heavy and refined petroleum products. The analytical performance, closed beam safety and ease of use of traditional bench top XRF units are all captured in theX-5000. And it delivers true field portability, being packaged into an easy to carry, 22 lb. battery operated XRF with integrated PC and industrialized large touch screen. |
![]() |
The X-5000 sets the benchmark for Performance, Power and Portability - No sample preparation, just collect it and analyze - right on the spot.
Key Features and Benefits
|
![]() |
The unique sample tray ensures proper placement of the sample every time for accurate analysis. The sample tray accommodates both bottles and cups of any size.
No sample preparation, no daily calibration.
The X-5000 is engineered to be used anywhere - in the field, at the production line, in an inspection area. It ensures operator safety as a fully interlocked, closed beam system. The closed beam, integrated design unique to the X-5000 is a critical safety advancement as portable XRF
analyzers continue to increase in x-ray power and be applied to ever more demanding, in-the-field analytical challenges. The X-5000 is your answer for the best analytical performance without compromising field portability or operator safety.
| Sample Type | ASTM D4294 | X-5000 |
| E10 | 52 | 54 |
| Diesel | 48 | 47 |
| Kerosene | 260 | 250 |
| B20 | 94 | 99 |
| E10 Gasoline | 96 | 101 |
| Jet A | 459 | 457 |
| No. 2 Diesel Oil | 878 | 858 |
| Hydraulic Fluid | 448 | 463 |
| Transformer Oil | 33 | 29 |
| Jet B5 | 696 | 712 |
| Mineral Oil | 29 | 30 |
| HFO NIST (2.996%) | 3.10% | 3.01% |
Table: Comparison D4294 to X-5000. Units are in mg/kg (ppm) except where indicated. X-5000 testing time was 120 seconds.

| Concentration Range | ppm to % levels |
| Analyzer Weight | 10 kg (22 lbs) |
| Measurement Time | 180 s |
| Operating Environment | -10 to 50 °C |
| Power Requirements | AC or Battery |
| Tube Voltage | 10 - 50 kV |
| Tube Current | 200 µA |
| User Interface | Built in Touchscreen |
| Instrument Dimensions |
38 x 33 28 cm
(15 x 13 x11 in.) |
| Sample Chamber |
29 x 11 x 15 cm
(11 x 6 x 5 in.) |
Das Röntgenfluoreszenzsystem X-5000 Mobile prüft mit energiedispersiver Röntgenfluoreszenz mit Laborqualität in einem tragbaren Gehäuse: auf Röntgenrohr basierend, geschlossenes Strahlungsgehäuse, eingebauter PC, großer Bildschirm und virtuelle Tastatur. 12,7 kg ohne Akku, ideal für Flüssigkeiten, Feststoffe oder Pulver, für tragbare Labs oder zur Prüfung von eingehenden/ausgehenden Stoffen auf der Produktionsstrasse.Copyright 2011 OLYMPUS CORPORATION, Alle Rechte vorbehalten. Nutzungsbedingungen | Datenschutzerklärung